Specifications
RIGOL Chapter 2 Command System
2-110 M300 Programming Guide
ROUTe:MONitor:STATe
Syntax ROUTe:MONitor:STATe <mode>
ROUTe:MONitor:STATe?
Description
Enable or disable the monitor function.
Parameters
Name
Type
Range
Default
<mode>
Bool
{OFF|0|ON|1}
OFF
Explanation In the monitor mode, the instrument performs continuous measurements on the
specified channels and displays the measurement values on the front panel display. At
most 7 channels can be monitored at the same time and you can change the channels
monitored at any time.
If no scan list is currently configured, the first channel of the multifunction module
(DIO channel) is monitored.
The channel monitor function can monitor the following channels: the multiplexer
channels (configured to the scan list and the DMM module is enabled), the DIO and
TOT channels of the multifunction module (no matter whether they are configured to
the scan list and the DMM module is not required).
Sending the command will generate
an error when the above conditions are not satisfied.
A scan (refer to the
ROUTe:SCAN command) always has priority over the monitor
function.
When the alarm function and scaling function are applied to the channel being
monitored, all the alarm data is stored in the alarm queue (which will be cleared at
power-off).
Readings acquired during a monitor are not stored in the memory but they are
displayed on the front panel screen, however, all the readings from a scan in progress
are stored in the memory.
Return
Format
The query returns 0 (OFF) or 1 (ON).
Example
CONF:VOLT:DC (@103:105)
ROUT:MON:CHAN (@103:105)
ROUT:MON:STAT ON
ROUT:MON:STAT?
The query returns 1.
Related
commands
ROUTe Command Subsystem
ROUTe:MONitor:DATA?
ROUTe:MONitor[:CHAN]