Specifications

Chapter 2 Command System RIGOL
M300 Programming Guide 2-109
ROUTe:MONitor:DATA?
Syntax
ROUTe:MONitor:DATA?
Description
Query the measurement values of all the monitored channels.
Explanation It returns the reading only; the units, time stamp, channel number and alarm status
information are not returned (not affected by the
FORMat Command Subsystem
commands).
If the monitor mode is disabled, an error indicating that it is unable to perform the
requested operation will be generated when sending this command.
Readings acquired during a monitor are not stored in the memory but they are
displayed on the front panel screen; however, all the readings from a scan in progress
are stored in the memory.
Return
Format
The query returns the readings of all the monitored channels in scientific notation.
Multiple return values are separated by commas.
Example ROUT:MON:DATA?
The query returns -6.514059579E-04,-1.011260443E-03.
Related
commands
ROUTe Command Subsystem
ROUTe:MONitor[:CHAN]
ROUTe:MONitor:STATe
ROUTe:MONitor:DATA:FULL?
Syntax
ROUTe:MONitor:DATA?
Description Query the scan readings of all the monitored channels (the readings are returned in full
format).
Explanation It returns the reading with the units, time stamp, channel number and alarm status
information (not affected by the
FORMat Command Subsystem commands).
If the monitor mode is disabled, an error indicating that it is unable to perform the
requested operation will be generated when sending this command.
Readings acquired during a monitor are not stored in the memory but they are
displayed on the front panel screen; however, all the readings from a scan in progress
are stored in the memory.
Return
Format
The query returns the readings of all the monitored channels in scientific notation.
Multiple return values are separated by commas.
Example ROUT:MON:DATA:FULL?
The query returns -1.538041765E-03V,2013,10,31,08,57,32.339,301,0.
Related
commands
ROUTe Command Subsystem
ROUTe:MONitor[:CHAN]