Specifications
RIGOL Chapter 2 Command System
2-102 M300 Programming Guide
ROUTe:CHANnel:ADVance:EDGE
Syntax ROUTe:CHANnel:ADVance:EDGE {RISing|FALLing}
ROUTe:CHANnel:ADVance:EDGE?
Description Set the instrument to open the current channel and close the next channel in the scan list
on the rising edge or falling edge of the input signal when the advanced source is set to
external (EXTernal). This command is only available when external digital multimeter is
used.
Parameters
Name
Type
Range
Default
{RISing|FALLing}
Discrete
RISing|FALLing
None
Explanation This command is valid only when the DMM module is disabled (refer to the
INSTrument:DMM command) or is not installed and it is also used to set the edge type
in the external trigger mode.
EXTernal: receive the external advanced channel signal input from the pin 6 of the
Alarms/Ext Trig interface (convert from the [RS-232/Alarms/Ext Trig] Interface).
The instrument opens the current channel and closes the next channel in the scan list
on the specified type od edge of the input signal.
As the “EXTernal” advanced source input and the external trigger input use the same
pin, the external trigger mode does not support “EXTernal” advanced source.
Alarms/Ext Trig Interface
Return
Format
The query returns the edge type currently specified (FALL (falling edge) or RIS (rising
edge)).
Example INST:DMM OFF
ROUT:SCAN (@202:220)
TRIG:SOUR IMM
TRIG:COUN 3
ROUT:CHAN:ADV:SOUR EXT
ROUT:CHAN:ADV:EDGE FALL
ROUT:CHAN:ADV:EDGE?
The query returns FALL.
Related
commands
ROUTe:CHANnel:ADVance:SOURce
INSTrument:DMM
Pin 6: Ext Trig Input/Channel
Advance Input