Specifications
RIGOL Chapter 2 Command System
2-100 M300 Programming Guide
ROUTe Command Subsystem
ROUTe:CHANnel:ADVance:SOURce
ROUTe:CHANnel:ADVance:EDGE
ROUTe:CHANnel:DELay
ROUTe:CHANnel:DELay:AUTO
ROUTe:CHANnel:FWIRe
ROUTe:CLOSe
ROUTe:CLOSe:EXCLusive
ROUTe:DONE?
ROUTe:MONitor[:CHAN]
ROUTe:MONitor:DATA?
ROUTe:MONitor:DATA:FULL?
ROUTe:MONitor:STATe
ROUTe:OPEN
ROUTe:SCAN
ROUTe:SCAN:SIZE?
ROUTe:SCAN:ADD
ROUTe:SCAN:REMove
ROUTe:CHANnel:ADVance:SOURce
Syntax ROUTe:CHANnel:ADVance:SOURce <source>
ROUTe:CHANnel:ADVance:SOURce?
Description Select the advanced source mode. This setting is only applicable when the external digital
multimeter is used. The advanced source provides advanced channel signals. The
instrument opens the current channel and close the next channel in the scan list when the
advanced channel signal is received.
Parameters
Name
Type
Range
Default
<source>
Discrete
{EXTernal|BUS|IMMediate}
EXTernal
Explanation This command is valid only when the DMM module is disabled (refer to the
INSTrument:DMM command) or is not installed.
EXTernal: receive the external advanced channel signal input from the pin 6 of the
Alarms/Ext Trig interface (convert from the [RS-232/Alarms/Ext Trig] Interface).
As the “EXTernal” advanced source input and the external trigger input use the same
pin, the external trigger mode does not support “EXTernal” advanced source.