Specifications

RIGOL
M300 Selection Guide 37
Appendix: Temperature Sensor Selection Reference
Temperature measurement is one of the main functions of M300. The temperature sensors used in the
temperature measurements are not the standard accessories or optional accessories of M300 and you
need to buy the temperature sensor according to the actual measurement need. This section lists the
recommended temperature sensor models for your reference.
1. Recommended Thermocouples
Recommended Model
Type
Temperature Range
Tolerance
[1]
Explanation
OMEGA 5TC-GG-J-20-72
OMEGA 5TC-GG-J-24-72
OMEGA 5TC-GG-J-30-72
OMEGA 5TC-TT-J-36-72
J
-40 to +375
±0.5℃
Can provide NIST
calibration
375 to 750
±0.004|t|
OMEGA 5TC-GG-K-20-72
OMEGA 5TC-GG-K-24-72
OMEGA 5TC-GG-K-30-72
OMEGA 5TC-TT-K-36-72
K
-40 to +375 ±1.5
375 to 1000
±0.004|t|
OMEGA 5TC-GG-T-20-72
OMEGA 5TC-GG-T-24-72
OMEGA 5TC-GG-T-30-72
OMEGA 5TC-TT-T-36-72
T
-40 to +125 ±1.5
125 to 350
±0.004|t|
OMEGA 5TC-GG-E-20-72
OMEGA 5TC-GG-E-24-72
OMEGA 5TC-GG-E-30-72
OMEGA 5TC-TT-E-36-72
E
-40 to +375 ±1.5
375 to 800
±0.004|t|
Note
[1]
: The specifications listed are the specifications of the thermocouples rather than those of M300.
2. Recommended RTDs
Recommended Model
Temperature Range
Temperature Precision
Explanation
OMEGA PR-10-2-100-1/8-6-E
OMEGA PR-10-2-100-1/8-12-E
OMEGA PR-10-2-100-1/8-18-E
OMEGA PR-10-2-100-1/8-24-E
OMEGA PR-10-2-100-1/4-6-E
OMEGA PR-10-2-100-1/4-12-E
OMEGA PR-10-2-100-1/4-18-E
OMEGA PR-10-2-100-1/4-24-E
-50
to 250
Conform to IEC CLASS B
±0.30@0
High-precision film
type 100Ω or 100Ω B
class DIN Platinum
element
Conform to IEC751
standard
α=0.00385Ω/Ω/
OMEGA PR-20-2-100-1/8-2-E-T
OMEGA PR-20-2-100-3/16-2-E-T
OMEGA PR-20-2-100-1/4-2-E-T
-50
to 260
Conform to IEC CLASS A
±0.15℃@ 0℃
High-precision 100Ω A
class DIN Platinum
element
Conform to IEC751
standard
α=0.00385Ω/Ω/
OMEGA
PR-20-2-100-1/8-2-E-G
OMEGA PR-20-2-100-3/16-2-E-G
OMEGA PR-20-2-100-1/4-2-E-G
-50
to 450