Instructions
RIGOL  Chapter 2 Performance Verification Test 
2-20   DSG800 Performance Verification Guide 
10 MHz Reference Input Amplitude Range Test 
Specification 
10 MHz Reference Input Amplitude Range 
Specification  0 dBm to +10 dBm 
Test Devices 
1.  Function/Arbitrary Waveform Generator × 1 
2.  Signal Analyzer × 1 
3.  Dual-BNC Cable × 2 
4.  N-BNC Cable × 1 
Test Connection Diagram 
10MHz OUT
DSG800
DG
4000
N9030A
CH1 CH2
10MHz IN
10MHz IN
RF IN
Dual-
BNC Cable
Dual-BNC Cable
N-BNC Cable
Figure 2-8 10 MHz Reference Input Amplitude Range Test Connection Diagram 
Test Procedures 
1.  Connect the [CH1] output terminal of the function/arbitrary waveform generator with the 
[10MHz IN] input terminal of the signal analyzer. Set the signal analyzer to external 
reference; set the frequency and amplitude of CH1 of the function/arbitrary waveform 
generator to 10 MHz and +5 dBm respectively and turn on the output to synchronize the 
function/arbitrary waveform generator and signal analyzer, as shown in Figure 2-8.  
Note: The 10 MHz signal output from the function/arbitrary waveform generator must have 
relatively higher accuracy (with lower than ±5 ppm deviation); otherwise, this specification 
cannot be measured. 