Instructions
RIGOL  Chapter 2 Performance Verification Test 
2-8   DSG800 Performance Verification Guide 
b)  Set the center frequency to 1 GHz. 
c)  Set the span to 50 kHz. 
d)  Set the resolution bandwidth to 1 kHz. 
e)  Set the video bandwidth to 3 Hz. 
f)  Set the reference level to 0 dBm. 
g)  Set the input attenuation to 10 dB. 
h)  Set the trace type to clear write. 
i)  Set the detector type to RMS average. 
j)  Set the sweep time to auto and the sweep time method to accurate. 
5.  Press Single, wait for the instrument to finish a sweep and press Peak Search to find the 
maximum peak.  
6.  Press Marker  Delta  input 20 kHz. Press Single and wait for the instrument to finish a 
sweep. Press Marker Function  Marker Noise, read the current measurement result and 
record it to the Test Record Form.  
7.  Press Mode Preset to reset the signal analyzer to the factory setting. Set the output 
frequency of DSG800 according to Table 2-3, repeat steps 4 to 6 (note that modify the center 
frequency of the signal analyzer accordingly in step 4 each time the output frequency of 
DSG800 is modified) and record the result.  
Table 2-3 Output Frequency of DSG800 
Output Frequency 
100 MHz  3 GHz (Only for DSG830) 
8.  Compare the measurement result with the specification. 
Test Record Form 
Single-sideband Phase Noise 
DSG800 Output 
Frequency 
Offset  Measurement Value  Limit  Pass/Fail 
100 MHz  Offset 20 kHz    < -100 dBc/Hz   
1 GHz  Offset 20 kHz    < -100 dBc/Hz   
3 GHz (Only for 
DSG830) 
Offset 20 kHz    < -94 dBc/Hz 