RIGOL  Chapter 2 Performance Verification Test 
2-2   DSG800 Performance Verification Guide 
Amplitude Accuracy Test 
Specifications 
Absolute Level Uncertainty 
Temperature range: 20℃ to 30℃ 
+13 dBm to -60 dBm 
-60 dBm to -110 dBm 
100 kHz ≤ f ≤ 3 GHz 
≤ 0.9 dB, 
≤ 0.5 (typ.) 
≤ 1.1 dB, 
≤ 0.7 (typ.) 
Test Devices 
1.  Power Meter and Power Sensor × 1 
2.  Dual-N Cable × 1 
3.  Dual-BNC Cable × 1 
4.  Signal Analyzer × 1 
Test Connection Diagram 
(a)
(b)
N9030A
Dual-N Cable
Dual-BNC Cable
RF IN
RF OUT
RF OUT
N1913A
E9304A
10MHz OUT
EXT REF IN
DSG800
DSG800
Figure 2-1 Amplitude Accuracy Test Connection Diagram 
Test Procedures 
1.  Calibrate the power meter: 
a)  Connect the power sensor with the [REF] terminal and channel A of the power meter. 
Press Channel and set the frequency of channel A to 50 MHz.  
b)  Press Cal and enable Power Ref in the Zero/Cal menu. Press Zero+Cal and wait for 