Datasheet
Electromagnetic Compatibility and Safety
Certicate of conformity
CE
cTUVus
EAC
EMC
Conform to EN61326-1:2013
IEC 61000-4-2:2008
±4.0 kV (contact discharge), ±8.0 kV (air
discharge)
IEC 61000-4-3:2006+A1+A2
3 V/m (80 MHz to 1 GHz)
3 V/m (1.4 GHz to 2 GHz)
1 V/m (2.0 GHz to 2.7 GHz)
IEC 61000-4-4:2004+A1 1 kV power cable
IEC 61000-4-5:2005
0.5 kV (Phase to Neutral)
0.5 kV (Phase to PE)
1 kV (Neutral to PE)
IEC 61000-4-6:2008 3 V, 0.15-80 MHz
IEC 61000-4-8:2009 3 A/m (50 Hz, 60 Hz)
IEC 61000-4-11:2004
Voltage dip:
0% UT during half cycle
0% UT during 1 cycle
70% UT during 25 cycles
Short interruption:
0% UT during 250 cycles
Safety regulation
Conform to:
UL 61010-1:2012
CAN/CSA-C22.2 No. 61010-1-12
EN 61010-1:2010
Environmental
Temperature
Operating temperature range 0
℃
to 50
℃
Storage temperature range -20
℃
to 70
℃
Humidity
0
℃
to 30
℃
≤ 95% rel. humidity
30
℃
to 40
℃
≤ 75% rel. humidity
40
℃
to 50
℃
≤ 45% rel. humidity
Altitude Operating height Below 3,000m
Dimensions
(W × H × D)
261.5 mm × 112 mm × 318.4 mm
(10.30 inch × 4.41 inch × 12.54 inch)
Weight
4.2 kg (9.3 Ib)
Calibration Interval
Recommended calibration interval 1 year