User Manual
DSA800 Series Specifications RIGOL
Page 6/17
Displayed Average Noise Level (DANL)
DSA832E
attenuation = 0 dB, RBW = VBW = 10 Hz, sample detector, trace
average ≥ 50, tracking generator off, 20℃ to 30℃, input
impendence = 50 Ω
PA off
9 kHz to 100 kHz
<-110 dBm (typ.)
100 kHz to 5 MHz
<-122 dBm, <-125 dBm (typ.)
5 MHz to 3.2 GHz
<-127 dBm, <-130 dBm (typ.)
PA on
100 kHz to 1 MHz
<-142 dBm (typ.)
1 MHz to 5 MHz
<-140 dBm, <-143 dBm (typ.)
5 MHz to 3.2 GHz
<-145 dBm, <-148 dBm (typ.)
Level Display
Logarithmic level axis
1 dB to 200 dB
Linear level axis
0 to reference level
Number of display points
601
Number of traces
3 + math trace
Trace detectors
normal, positive-peak, negative-peak, sample, RMS, voltage
average
quasi-peak (with EMI-DSA800 option)
Trace functions
clear write, max hold, min hold, average, view, blank
Units of level axis
dBm, dBmV, dBμV, nV, μV, mV, V, nW, μW, mW, W