User Manual

DSA800 Series Specifications RIGOL
Page 15/17
Mass Memory
Mass memory
flash disk (internal), USB storage device (not supplied)
Power Supply
Input voltage range, AC
100 V to 240 V (nom.)
AC supply frequency
45 Hz to 440 Hz
Power consumption
35 W (typ.), max. 50 W with all options
Environmental
Temperature
operating
temperature range
0 to 50
storage temperature
range
-20 to 70
Humidity
0 to 30
95% rel. humidity
30 to 40
75% rel. humidity
Altitude
operating height
up to 3,000m
Electromagnetic Compatibility and Safety
EMC
in line with EN61326-1:2006
IEC 61000-4-2:2001
± 4.0 kV (contact discharge), ± 4.0 kV (air discharge)
IEC 61000-4-3:2002
3 V/m (80 MHz to 1 GHz), 3 V/m (1.4 GHz to 2 GHz), 1
V/m (2.0 GHz to 2.7 GHz)
IEC 61000-4-4:2004
1 kV power lines
IEC 61000-4-5:2001
0.5 kV (phase to neutral), 0.5 kV (phase to PE), 1 kV
(neutral to PE)
IEC 61000-4-6:2003
3 V, 0.15 to 80 MHz
IEC 61000-4-11:2004
voltage dip: 0% UT during half cycle, 0% UT during 1
cycle, 70% UT during 25 cycles
short interruption: 0% UT during 250 cycles
Electrical safety
in line with
UL 61010-1:2012, CAN/CSA-C22.2 No. 61010-1-12, EN
61010-1:2010
Dimensions
(W x H x D)
361.6 mm × 178.8 mm × 128 mm
(14.2 in × 7.0 in × 5.0 in)
Weight
DSA815
DSA832E
DSA832
DSA875
Standard
4.25 kg (9.4 lb)
4.55 kg (10.0 lb)
With tracking generator
5.15 kg (11.4 lb)
Calibration Interval
Recommended calibration interval
1 year