User Manual
DSA800 Series Specifications RIGOL
Page 11/17
Distortion
Second Harmonic Intercept
DSA815
DSA832E
DSA832
DSA875
Second harmonic
intercept (SHI)
f
c
≥ 50 MHz, input signal level = -20 dBm, attenuation = 10 dB
+40 dBm
+40 dBm
+45 dBm
Third-order Intercept
DSA815
DSA832E
DSA832
DSA875
Third-order intercept
(TOI)
f
c
≥ 50 MHz, two -20 dBm tones at input mixer spaced by 200 kHz,
attenuation = 10 dB
+10 dBm
+7 dBm
+11 dBm, +15 dBm (typ.)
1dB Gain Compression
1dB compression of input
mixer (P
1dB
)
f
c
≥ 50 MHz, attenuation = 0 dB
>0 dBm