Instructions
RIGOL  Chapter 2 Front Panel Operation 
User’s Guide for DSA800/E Series 
2-30 
Enter contin
uous s
weep mode and 
w
ait for trigger 
condition
Sweep 
again and wait for t
rigger 
c
ondition
Select 
“
Cont
”
Single sweep status?
Y
N
Figure 2-9 Process of Continuous Sweep 
Single 
In single sweep mode, this menu is used to execute trigger initialization. After that, 
the analyzer will execute the specified number of sweeps (or measurements) if the 
trigger conditions are satisfied. 
Key Points: 
  If the instrument is in continuous sweep mode and no measurement function 
is enabled, press this key and the system will enter single sweep mode and 
execute the specified number of sweeps if the trigger conditions are satisfied.     
  If the instrument is in continuous sweep mode and a measurement function is 
on, press this key and the system will enter single measurement mode and 
execute the specified number of measurements if the trigger conditions are 
satisfied. 
  If the system is already in single sweep mode, press this key and the system 
will execute the specified number of sweeps (or measurements) if the trigger 
conditions are satisfied. 
  In single sweep mode, trigger initialization (press Sweep/Trig  Single or 
send the “:INIT” command through the remote interface) should be executed 
before trigger condition judgment.  