Instructions
Chapter 2 Front Panel Operation  RIGOL 
User’s Guide for DSA800/E Series   
2-21 
Table 2-11 Edit menus of amplitude correction 
Menu  Explanation 
Point  Create or edit the data point of a correction factor. 
Range: 1 to 200 
Frequency 
Set the frequency of the specified point in the correction 
factor. 
Amplitude 
Set the amplitude correction of the specified point in the 
correction factor. 
Range: -120 dB to 100 dB 
Del Point 
Delete the specified point data in the correction factor: 
frequency and amplitude correction.  
4.  Freq Interp 
Set the interpolation type of the points between two points in the correction 
table during amplitude correction. 
  In Lin mode, the frequency and amplitude use a Lin and Log unit to 
perform interpolation respectively. 
  In Log mode, both of the parameters use a Log unit. 
5.  Delete 
Clear all the frequency and amplitude correction data of the selected 
correction factor. 
6.  Corr Table 
Turn on the correction table to view the correction data edited. At this 
moment, the screen is divided into two parts with the upper showing the 
measurement curve and the lower showing the points edited, frequency and 
amplitude. 
7.  Corr View 
  All: view the data of all the correction factors.   
  Sel: view the data of the selected correction factor.   