Instructions
Chapter 2 Front Panel Operation  RIGOL 
User’s Guide for DSA800/E Series   
2-7 
Start
Execute a 
sweep
Search for 
peak and 
mark it as 
Marker
Active Marker 
exists?
N
Set marker 
frequency to 
center freq
Y
Execute the 
next sweep
Search for and mark the 
frequency point (variation 
< 3dB) near the Marker 
Figure 2-1 Process of Signal Track 
Key Points: 
  When Signal Track is turned on, the ST (Signal Track) icon 
  is highlighted 
in the status bar at the left of the screen. 
  If an active marker currently exists, when Signal Track is enabled, the 
instrument will search for and mark the point (with no more than 3 dB 
variation in amplitude) near the marker as well as set the frequency of this 
point as the center frequency to hold the signal at the center of the screen. 
  If no marker is currently active, when Signal Track is enabled, the instrument 
will activate Marker 1, execute a peak search automatically and set the 
frequency of the current peak as the center frequency to hold the signal at the 
center of the screen. 
  In continuous sweep, the system tracks the signal continuously; in single 
sweep, the instrument only performs a single signal track; in Zero Span, 
Signal Track is invalid.    