Instructions
RIGOL  Chapter 5 Specifications 
User’s Guide for DSA800/E Series 
5-8 
Absolute Amplitude Uncertainty 
  DSA815  DSA832  DSA875  DSA832E 
Uncertainty 
f
c
 = 50 MHz, peak detector, preamplifier off, 
attenuation = 10 dB, input signal level = -10dBm, 
20℃  to 30℃ 
<0.4 dB  <0.3 dB 
RBW Switching Uncertainty 
Uncertainty 
relative to 1 kHz RBW  
<0.1 dB 
Reference Level 
Range  -100 dBm to +20 dBm, in 1 dB step 
Resolution 
log scale  0.01 dB 
linear scale  4 digits 
Preamplifier 
  DSA815  DSA832  DSA875  DSA832E 
Gain 
100 kHz to 1.5 GHz 
20 dB 
(nom.) 
17 dB 
(nom.) 
17 dB 
(nom.) 
17 dB 
(nom.) 
1.5 GHz to 3.2 GHz   
3.2 GHz to 7.5 GHz       
Level Measurement Uncertainty 
DSA815  DSA832  DSA875  DSA832E 
95% confidence level, S/N > 20 dB, RBW = VBW = 1 
kHz, preamplifier off, attenuation = 10 dB, -50 dBm 
< input level ≤ 0 dBm, f
c 
> 10 MHz, 20℃  to 30℃ 
Level measurement 
uncertainty 
<1.5 dB 
(nom.) 
<0.8 dB (nom.) 
<1.0 dB 
(nom.) 