Data Sheet
RIGOL
5
Specications
Frequency
Frequency
DSA815 DSA832 DSA875 DSA832E
Frequency range 9 kHz to 1.5 GHz 9 kHz to 3.2 GHz 9 kHz to 7.5 GHz 9 kHz to 3.2 GHz
Frequency resolution 1 Hz
Internal Reference Frequency
DSA815 DSA832 DSA875 DSA832E
Reference frequency 10 MHz
Accuracy
± [(time since last adjustment × aging rate) + temperature stability +
calibration accuracy]
Initial calibration accuracy <1 ppm
Temperature stability
0
℃
to 50
℃
, reference to 25
℃
<2 ppm <0.5 ppm
<1 ppm
Aging rate <2 ppm/year <1 ppm/year
<2 ppm/year
Frequency Readout Accuracy
Marker resolution span/ (number of sweep points - 1)
Marker uncertainty
± (frequency indication × frequency reference uncertainty + 1% × span + 10% × resolution
bandwidth + marker resolution)
Frequency Counter
Resolution 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz
Uncertainty ± (frequency indication × reference frequency accuracy + counter resolution)
Frequency Span
Range 0 Hz, 100 Hz to maximum frequency of instrument
Uncertainty ±span/ (number of sweep points - 1)
SSB Phase Noise
20
℃
to 30
℃
, f
c
=1 GHz
Carrier offset DSA815 DSA832 DSA875 DSA832E
10 kHz <-80 dBc/Hz <-98 dBc/Hz
<-90 dBc/Hz,
<-98 dBc/Hz (typ.)
100 kHz <-100 dBc/Hz (typ.) <-100 dBc/Hz (typ.) <-100 dBc/Hz (typ.)
Specications are valid under the following conditions: the instrument is within the calibration period, is stored for at least two hours at 0
℃
to 50
℃
temperature, and is warmed up for 40 minutes. Unless otherwise noted, the specications in this manual include the measurement
uncertainty.
Typical (typ.):
characteristic performance, which 80 percent of the measurement results will meet at room temperature (approximately 25
℃
). This data is not
warranted and does not include the measurement uncertainty.
Nominal (nom.):
the expected mean or average performance or a designed attribute (such as the 50 Ω connector). This data is not warranted and is measured at
room temperature (approximately 25
℃
).
Measured (meas.):
an attribute measured during the design phase which can be compared to the expected performance, such as the amplitude drift variation with
time. This data is not warranted and is measured at room temperature (approximately 25
℃
).
NOTE: All charts in this manual are the measurement results of multiple instruments at room temperature unless otherwise noted. The specications (except the
TG specications) listed in this manual are those when the tracking generator is off.