Calibration Interval
Recommended calibration interval 18 months
Dimensions
(W x H x D)
361.6 mm × 178.8 mm × 128 mm
(14.2 in × 7.0 in × 5.0 in)
Weight
DSA815 DSA832 DSA875 DSA832E
Standard
4.25 kg (9.4 lb)
4.55 kg (10.0 lb)
With tracking generator 5.15 kg (11.4 lb)
Electromagnetic Compatibility and Safety
EMC
in line with EMC instruction (2014/30/EU),
in line with or exceed IEC61326-1: 2013/EN61326-1: 2013 Group 1 Class A standard
CISPR 11/EN 55011
IEC 61000-4-2:2008/EN 61000-4-2 ±4.0 kV (contact discharge), ±8.0 kV (air discharge)
IEC 61000-4-3:2002/EN 61000-4-3
3 V/m (80 MHz to 1 GHz); 3 V/m (1.4 GHz to 2 GHz);
1 V/m (2.0 GHz to 2.7 GHz)
IEC 61000-4-4:2004/EN 61000-4-4 1 kV power lines
IEC 61000-4-5:2001/EN 61000-4-5 0.5 kV (phase to neutral); 1 kV (phase to PE); 1 kV (neutral to PE)
IEC 61000-4-6:2003/EN 61000-4-6 3 V, 0.15-80MHz
IEC 61000-4-11:2004/EN 61000-4-11
voltage dip: 0% UT during half cycle; 0% UT during 1 cycle; 70% UT during 25
cycles short interruption: 0% UT during 250 cycles
Electrical safety
IEC 61010-1:2010 (Third Edition)/EN 61010-1:2010,
UL 61010-1:2012 R4.16 and CAN/CSA-C22.2 NO. 61010-1-12+ GI1+ GI2