2. Avg Mode
Set the average operation mode to Exp or Repeat and the default is Exp.
When Exp is selected, the result is the exponential average of the
current N (N is specified in Avg Num) measurement results.
When Repeat is selected, the result is the arithmetic average of the
current N (N is specified in Avg Num) measurement results.
3. TP Type
Peak
Display the power of the signal with the maximum amplitude between the
start line and stop line. The detector type is set to Pos Peak
automatically.
Average
Display the average power of signals between the start line and stop line.
The detector type is set to Voltage Avg automatically.
RMS
Display the mean square root (in power unit) of the voltages of signals
between the start line and stop line. The detector type is set to RMS
Avg automatically.
4. Start Line
Set the left margin (in time unit) of T-Power measurement. The data
calculated under this measurement is between the start line and stop line.
You can use the numeric keys, knob or direction keys to modify this
parameter.
Table 2-28 Start Line for T-Power Measurement
Parameter
Description
Default
0 us
Range
0 us to Stop Line
Unit
Ks, s, ms, us, ns, ps
Knob Step
Sweep Time/600
Direction Key Step
in 1-1.5-2-3-5-7.5 sequence