User manual

User’s Guide for DSA800 Series
Chapter 2 Front Panel Operation RIGOL
2-43
Table 2-23 Power Range of TG
Parameter
Explanation
Default
0 dB
Range
0 dB to 20 dB
Unit
dB
Knob Step
1 dB
Direction Key Step
10 dB
Normalize
Eliminate the error of TG Level. Before using this function, connect the [GEN
OUTPUT 50 Ω] output terminal of the TG with the [RF INPUT 50 Ω] input
terminal of the analyzer.
1. Stor Ref
Save the data of Trace 1 to Trace 3 as the reference value for normalization.
This operation should be done before you enable the normalization.
2. Normalize
Enable or disable the normalization. When enabled, the reference trace will be
stored automatically after the current sweep finishes if no reference trace is
stored before. During storing the reference trace, the corresponding prompt
message is displayed. When normalization is enabled, the corresponding
value of the reference trace will be substracted from the trace data after every
sweep.
3. Norm Ref Lvl
Adjust the vertical position of the trace on the screen when normalization is
enabled.
Being different from the Ref Level function in the AMPT menu, this
parameter has no influence on the reference level of the analyzer.
You can use the numeric keys, knob or direction keys to modify this
parameter. For more details, please refer to Parameter Setting.