RIGOL Chapter 2 Front Panel Operation
User’s Guide for DSA800 Series
2-30
Enter continuous sweep mode and
wait for trigger condition
Sweep again and wait for trigger
condition
Select Cont
Single sweep status
Y N
Figure 2-9 Process of Continuous Sweep
Single
In single sweep mode, this menu is used to execute trigger initialization. After that,
the analyzer will execute specified number of sweeps (or measurements) if all the
trigger conditions are satisfied.
Key Points:
If the instrument is in continuous sweep mode and no measurement function
is enabled, press this key and the system will enter the single sweep mode
and execute specified number of sweeps if all the trigger conditions are
satisfied.
If the instrument is in continuous sweep mode and a measurement function is
on, press this key and the system will enter the single measure mode and
execute specified number of measurements if all the trigger conditions are
satisfied.
If the system is already in single sweep mode, press this key and the system
will execute specified number of sweeps (or measurements) if all the trigger
conditions are satisfied.
In single sweep mode, trigger initialization (press Sweep/Trig Single or
send the :INIT command through the remote interface) should be executed
before trigger condition judgment.