User manual

RIGOL Chapter 2 Front Panel Operation
User’s Guide for DSA800 Series
2-16
Volts in Lin scale type when the scale type is changed from Log to Lin.
The scale type does not affect the unit of Y-axis.
Units
Set the unit of the Y-axis to dBm, dBmV, dBuV, Volts or Watts. Wherein, dBm,
dBmV, and dBuV are for Log scale; Volts and Watts are for Linear scale. The
default is dBm.
Key Points:
The conversion relationships between units:
WR
Volts
B
001.0
1
log10md
2
(2-4)
V
Volts
VB
1
10
log20d
6
(2-5)
V
Volts
VB
m1
10
log20md
3
(2-6)
R
Volts
Watts
2
(2-7)
Wherein,
R
denotes the reference resistance.
Ref Offset
Assign an offset to the reference level to compensate for gains or losses generated
between the device under measurement and the analyzer.
Key Points:
The change of this value changes both the reference level readout and the
amplitude readout of the marker, but does not impact the position of the curve