Chapter 2 Front Panel Operation RIGOL
2-15
Scale/Div
Set the logarithmic units per vertical grid division on the display. This function is
only available when the scale type is set to “Log”.
Key Points:
By changing the scale, the amplitude range available is adjusted.
The range of the amplitude that can be displayed:
Minimum: reference level – 10 × the current scale value
Maximum: the reference level.
You can modify this parameter using the numeric keys, knob or direction keys.
For more details, please refer to “Parameter Setting”.
Table 2-8 Scale
Parameter
Explanation
Default
10 dB
Range
0.1 dB to 20 dB
Unit
dB
Knob Step
Scale ≥ 1, step = 1 dB
Scale < 1, step = 0.1 dB
Direction Key Step
in 1, 2, 5 sequence
Scale Type
Set the scale type of Y-axis to Lin or Log, the default is Log.
Key Points:
In Log scale type: the Y-axis denotes the logarithmic coordinate, the value
shown at the top of the grid is the reference level and each grid represents the
scale value. The unit of Y-axis will automatically switch to the default “dBm” in
Log scale type when the scale type is changed from Lin to Log.
In Lin scale type: the Y-axis denotes the linear coordinate, the values shown
at the top of the grid and the bottom of the grid are the reference level and 0
V respectively. Each grid represents 10% of the reference level and the
Scale/Div is invalid. The unit of Y-axis will automatically switch to the default