User manual
User’s Guide for DSA800 Series
Chapter 2 Front Panel Operation RIGOL
2-5
Table 2-3 Stop frequency
Parameter
Explanation
Default
1.5 GHz
Range
*
0 Hz to 1.5 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
Span > 0, step = Span/200
Span = 0, step = VBW/100
Min = 1 Hz
Direction Key Step
CF step
*
Note: The range is from 100 Hz to 1.5 GHz in non-zero span.
CF Step
Set the step of center frequency. Changing the center frequency in a fixed step
continuously switches the channel to be measured.
Key Points:
The CF step can be set in “Manual” or “Auto” mode. In Auto mode, the CF step
is 1/10 of the span in Non-zero span mode or equals the RBW while in Zero
span mode; in Manual mode, you can set the step using the numeric keys.
After you set an appropriate CF step and select Center Freq, use up and
down direction keys to switch between measurement channels in a specified
step in order to sweep the adjacent channel manually.
You can modify this parameter using the numeric keys, knob or direction keys.
For more details, please refer to “Parameter Setting”.
Table 2-4 CF step
Parameter
Explanation
Default
150 MHz
Range
1 Hz to 1.5 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
Span > 0, step = Span/200
Span = 0 , step = 100 Hz
Min = 1 Hz
Direction Key Step
in 1, 2, 5 sequence