User manual

User’s Guide for DSA800 Series
Chapter 2 Front Panel Operation RIGOL
2-91
Min Search
Search and mark the peak with the minimum amplitude on the trace.
Peak Peak
Execute peak search and minimum search at the samt time and mark the results
with a pair of markers. Wherein, the result of peak search is marked with the delta
marker and the result of minimum search is marked with the reference marker.
Cont Peak
Enable or disable continuous peak search and the default is Off. When enabled, the
system will always execute a peak search automatically after each sweep in order
to track the signal under measurement.
The difference between Cont Peak and Signal Track
In Cont peak, the system always searches for the maximum in the current channel,
while in Signal Track, the system always tracks the signal with the same amplitude
as the marker before Signal Track was enabled and uses the frequency of this signal
as the center frequency.
Search Para
Define the conditions of peak search for various peak searches. A real peak should
meet the requirements of both the PK Excursn and PK Thresh.
1 PK Excursn
Set the excursion between the peak and the minimum amplitude on both
sides of it. Peaks whose excursions are beyond the specified excursion are
treated as real peaks.