User manual
RIGOL Chapter 2 Front Panel Operation
User’s Guide for DSA800 Series
2-28
Sweep/Trig
Set parameters about the Sweep and Trigger functions, including Time, Auto SWT,
Mode, Numbers, Trigger type, etc.
Sweep Time
Set the time needed for the spectrum analyzer to finish a sweep within the span
range. The sweep time can be set in “Auto” or “Manual” mode and the default is
“Auto”.
Key Points:
In non-zero span, the analyzer selects the shortest sweep time on the basis of
the current RBW and VBW settings if Auto is selected.
Decreasing the sweep time would speed the measurement. However, an error
may be caused if the specified sweep time is less than the minimum sweep
time in Auto coupling, and “UNCAL” is shown in the status bar on the screen.
You can modify this parameter using the numeric keys, knob or direction keys.
For more details, please refer to “Parameter Setting”.
Table 2-15 Sweep time
Parameter
Explanation
Default
50 ms
Range
*
20 us to 1500 s
Unit
ks, s, ms, us, ns, ps
Knob Step
Sweep time/100, Min = 1 ms
Direction Key Step
in 1, 1.5, 2, 3, 5, 7.5 sequence
*
Note: The minimum in non-zero span mode is 10 ms.