Normalize 
Nomalization can eliminate the error of TG Level. Before using this function, 
connect the [GEN OUTPUT 50Ω] output terminal of the TG with the [RF INPUT 
50Ω] input terminal of the analyzer.  
1.  Stor Ref 
Save the data of Trace 1 to Trace 3 as the reference value for normalization. 
This operation should be done before you enable the normalization.  
2.  Normalize 
Enable or disable the normalization. When enabled, the reference trace will be 
stored automatically after the current sweep finishes if no reference trace is 
stored before. During the reference trace storage, the corresponding prompt 
message is displayed. When normalization is enabled, the corresponding 
value of the reference trace will be substracted from the trace data after every 
sweep.  
3.  Norm Ref Lvl 
Adjust the vertical position of the trace on the screen by adjusting the 
reference level when normalization is enabled. 
  Being different from the Ref Level function in the AMPT menu, this 
parameter has no influence on the reference level of the analyzer. 
  You can use the numeric keys, knob or direction keys to modify this 
parameter. For more details, please refer to “Parameter Setting”.  
Table 2-25 Reference level under normalization 
Parameter  Explanation 
Default  0 dB 
Range 
-200 dB to 200 dB 
Unit 
dB 
Knob Step  1 dB 
Direction Key Step  10 dB 
4.  Norm Ref Pos   
Adjust the vertical position of the normalization reference level on the screen 