User Manual
RIGOL  Chapter 2 Front Panel Operation 
User’s Guide for DSA800 Series 
2-38 
Result:   
The results of Pass/Fail test contain the pass number and the pass ratio. 
1.  Switch 
Turn on or off the Pass/Fail test function and the default is “Off”. When the 
function is enabled, the system changes to split-screen display. At this point, 
T2 and T3 are used for marking the upper and lower limits separately. 
2.  Settings 
  Limit: select the desired limit line (upper or lower) for editing. Then, 
press Edit to set the limit conditions. 
  Test: enable or disable the test function. Each limit line has its own 
switch. You can test according to both together, either one or neither. 
Note: Trace/P/F  Pass/Fail  Switch is only used to enable or 
disable the setting menu of Pass/Fail test, not the test itself.  
  Edit: edit the properties of the limit lines. 
Note: The edited limit line data can be stored in the internal or external 
memory of the analyzer and can be recalled when needed. After the 
editing is finished, press Storage to store the data according to the 
method introduced in “Storage”. 
Table 2-21 Pass/Fail menu 
Function  Explanation 
Limit  Quickly select the limit line to be edited.  
Point  Set the number of the point to be edited. The range is from 
1 to 200. 
X-axis  Edit the X-axis value (frequency or time) of the current 
point. If the X-axis unit is frequency and the Rel Freq is 
enabled, edit the frequency difference between the 
frequency of the current point and the center frequency. 
Amplitude  Edit the amplitude of the current point. If the Rel Ampt is 
enabled, edit the amplitude difference between the 
amplitude of the current point and the reference level. 
Connected  Connect this point with the previous one or not. 
Del Point
Delete the point you are editing. 










