User Manual
Chapter 2 Front Panel Operation  RIGOL 
User’s Guide for DSA800 Series   
2-23 
Sweep and Function Settings 
BW/Det 
Set the RBW (Resolution Bandwidth), VBW (Video Bandwidth) and detector type 
parameters of the analyzer. 
RBW 
Set the desired resolution bandwidth in order to distinguish between signals which 
are close in frequency. 
Key Points: 
  Reducing RBW will increase the frequency resolution, but will increase the 
sweep time (Sweep Time is affected by a combination of RBW and VBW when 
it is in Auto mode). 
  RBW varies with the span (non-zero span) in Auto RBW mode.  
  You can modify this parameter using the numeric keys, knob or direction keys. 
For more details, please refer to “Parameter Setting”.  
Note: If the detector type is “Quasi-Peak” or the filter type is “EMI”, RBW can be 
200 Hz, 9 kHz or 120 kHz only. 
Table 2-13 RBW (Filter Type is Gauss) 
Parameter  Explanation 
Default  1 MHz 
Range 
10 Hz to 1 MHz 
Unit 
GHz, MHz, kHz, Hz 
Knob Step  in 1, 3, 10 sequence 
Direction Key Step  in 1, 3, 10 sequence 