User Manual
RIGOL  Chapter 2 Front Panel Operation 
User’s Guide for DSA800 Series 
2-20 
Correction   
Correct the amplitude in order to compensate for the gain or loss from external 
devices such as Antenna and Cable. When using this function, you can view the 
correction data table and save or load the current correction data. When amplitude 
correction is turned on, both the trace and related measurement results will be 
corrected; the corresponding icon 
 is highlighted in the status bar at the left of 
the screen. 
1.  Select
Select a correction factor from Antenna, Cable, Other and User for the current 
correction and the default is Off (turn off all the correction factors). After 
chooosing the desired correction factors, press Correction to enable the 
correction factors selected. Multiple correction factors can be enabled at the 
same time.  
2.  Correction 
Enable or disable amplitude correction and the default is Off.  
When amplitude correction is enabled, the data of the correction factor 
currently selected is used for amplitude correction. If multiple factors are 
enabled, all related data will be used for amplitude correction. 
3.  Edit 
Edit the frequency of the correction factor and the correction data of the 
corresponding amplitude. You can modify this parameter using the numeric 
keys, knob or direction keys. For details, refer to the table on the next page. 
Note: The edited correction data can be stored in the internal or external 
memory of the analyzer and can be recalled when needed. When correction 
data editing is finished, press Storage to save the correction data using the 
method introduced in “Storage”. You can increase the number of point to 2 
after “point 1” is edited.  