User Manual
Chapter 2 Front Panel Operation  RIGOL 
User’s Guide for DSA800 Series   
2-13 
Figure 2-5 After Auto Scale 
Ref Level   
Set the maximum power or voltage can be currently displayed in the window. The 
value is displayed at the upper left corner of the screen grid.  
Key Points: 
  The maximum reference level available is affected by the maximum mixing 
level, input attenuation and preamplifier. When you adjust it, the input 
attenuation is adjusted under a constant maximum mixing level in order to 
fulfill the following inequality: 
mixPARFRef
L
aaL ≤
+−
                     (2-3) 
Wherein, 
Ref
L
, 
RF
a
, 
PA
a
 and 
mix
L
 denote the reference level, input 
attenuation, preamplifier and maximum mixing level respectively. 
  You can modify this parameter using the numeric keys, knob or direction keys. 
For more details, please refer to “Parameter Setting”.  