User Manual
Chapter 2 Front Panel Operation  RIGOL 
User’s Guide for DSA800 Series   
2-5 
Table 2-3 Stop Frequency 
Parameter  Explanation 
Default  7.5 GHz 
Range*  0 Hz to 7.5 GHz 
Unit  GHz, MHz, kHz, Hz 
Knob Step  Span > 0, step = Span/200 
Span = 0, step = VBW/100 
Min = 1 Hz 
Direction Key Step 
CF step 
Note: *The range is from 100 Hz to 7.5 GHz in non-zero span.  
Freq Offset 
You can set a frequency offset value to account for frequency conversions between 
the device under test and the input terminal of the spectrum analyzer. 
Key Points: 
  The change of this parameter only changes the display values of the center 
frequency, start frequency and stop frequency; but does not impact any 
hardware settings of the spectrum analyzer.  
  You can modify this parameter using the numeric keys, knob or direction keys. 
For more details, please refer to “Parameter Setting”. 
  To eliminate an offset, you can perform Preset operation or set the frequency 
offset to 0Hz. 
Table 2-4 Frequency Offset 
Parameter  Explanation 
Default  0 Hz 
Range  -100 GHz ~ 100 GHz 
Unit  GHz, MHz, kHz, Hz 
Knob Step  37.5 MHz 
Direction Key Step  CF Step 