User Manual
RIGOL  Chapter 5 Specifications 
User’s Guide for DSA800 Series 
5-6 
Displayed Average Noise Level (DANL) (Normalized to 1Hz) 
  DSA815  DSA832  DSA875  
Frequency 
attenuation = 0 dB, RBW = VBW = 100 Hz, sample 
detector, trace average ≥ 50, tracking generator off, 
normalized to 1Hz, 20℃ to 30℃, input impendence = 50 Ω 
PA 
off 
9 kHz to 100 kHz    <-120 dBm (typ.)  <-120 dBm (typ.) 
100 kHz to 1 MHz 
<-110 dBm,  
<-130 dBm (typ.) 
<-135 dBm, 
<-138 dBm (typ.) 
<-135 dBm, 
<-138 dBm (typ.) 
1 MHz to 5 MHz  <-130 dBm + 6 × 
(f/1 GHz) dB,  
<-135 dBm (typ.) 
5 MHz to 1.5 GHz 
<-140 dBm, 
<-144 dBm (typ.) 
<-140 dBm, 
<-144 dBm (typ.) 
1.5 GHz to 3.2 
GHz 
3.2 GHz to 6 GHz     
<-136 dBm, 
<-140 dBm (typ.) 
6 GHz to 7.5 GHz 
<-131 dBm, 
<-135 dBm (typ.) 
PA 
on 
100 kHz to 1 MHz 
<-130 dBm,  
<-150 dBm (typ.) 
<-152 dBm (typ.)  <-152 dBm (typ.) 
1 MHz to 5 MHz 
<-150 dBm + 6 × 
(f/1 GHz) dB, 
<-155 dBm (typ.) 
<-152 dBm, 
<-155 dBm (typ.) 
<-152 dBm, 
<-155 dBm (typ.) 
5 MHz to 1.5 GHz 
<-157 dBm, 
<-161 dBm (typ.) 
<-157 dBm,  
<-161 dBm (typ.) 
1.5 GHz to 3.2 
GHz 
3.2 GHz to 6 GHz     
<-153 dBm,  
<-157 dBm (typ.) 
6 GHz to 7.5 GHz     
<-148 dBm,  
<-152 dBm (typ.) 