Marker Fctn 
Special marker functions including Noise Mkr, N dB BW and Freq Count. 
Select Mkr 
Select the marker to be used for the specified measurement function and the 
default is marker 1. 
Noise Mkr 
Execute the Noise marker function for the selected marker and read the noise 
power spectral density.  
Key Points: 
  If the current marker is “Off” in the Marker menu, pressing Noise Mkr will 
first set it to Normal type automatically; then measure the average noise level 
at the marked point and normalize this value to 1 Hz bandwidth. During this 
process, certain compensation is always made on the basis of the detection 
and trace types. The measurement will be more precise if RMS Avg or Sample 
detection type is used. 
  This function can be used for measuring the C/N ratio. 
N dB BW 
Enable the N dB BW measurement or set the value of N dB.  
The N dB BW denotes the frequency difference between two points that are 
located on both sides of the current marker and with N dB fall (N<0) or rise (N>0) 
in amplitude as shown in the figure on the next page.  