Chapter 2 Front Panel Operation  RIGOL 
User’s Guide for DSA800 Series   
2-69 
Table 2-44 Average Number for C/N Ratio Measurement  
Parameter  Description 
Default  10 
Range  1 to 1000 
Unit  N/A 
Knob Step  1 
Direction Key Step  1 
2.  Avg Mode 
Set the average operation mode to “Exp” or “Repeat” and the default is “Exp”. 
  When “Exp” is selected, the result is the exponential average of the 
current N (N is specified in “Avg Num”) measurement results.   
  When “Repeat” is selected, the result is the arithmetic average of the 
current N (N is specified in “Avg Num”) measurement results.  
3.  Offset Freq 
Set the difference between the carrier center frequency and the noise center 
frequency. You can use the numeric keys, knob and direction keys to modify 
this parameter.  
Table 2-45 Frequency Offset for C/N Ratio Measurement 
Parameter  Description 
Default  2 MHz 
Range  33 Hz to 2.5 GHz 
Unit 
GHz, MHz, kHz, Hz 
Knob Step  Offset Frequency/100, the 
minimum is 1Hz 
Direction Key Step  in 1-1.5-2-3-5-7.5 sequence 
4.  Noise BW 
Set the bandwidth of the noise to be measured. You can use the numeric keys, 
knob or direction keys to modify this parameter.  