User Manual
RIGOL  Chapter 2 Front Panel Operation 
User’s Guide for DSA800 Series 
2-68 
C/N Ratio 
Measurement Interface: 
Figure 2-19 C/N Ratio Measurement Interface 
Measurement Results: carrier power, noise power and C/N ratio.  
  Carrier Power: power within the carrier bandwidth. 
  Noise Power: power within the noise bandwidth. 
  C/N Ratio: the ratio of the carrier power to the noise power. 
Measurement Parameters: average number, average mode, offset frequency, 
noise bandwidth and carrier bandwidth. 
1.  Avg Num 
Specify the number of averages used when calculating the measurement 
result and the default is “Off”. You can use the numeric keys, knob or direction 
keys to modify this parameter.  
Offset Frequency 
Carrier Bandwidth 
Noise Bandwidth 