User Manual
Chapter 2 Front Panel Operation  RIGOL 
User’s Guide for DSA800 Series   
2-65 
EBW 
Measurement Interface: 
Figure 2-18 EBW Measurement Interface 
Measurement Results: emission bandwidth, namely the bandwidth between 
two points on a signal which are X dB below the highest point within the span. 
During the measurement, the analyzer first determines the frequency (f
0
) of the 
maximum amplitude point within the span and then finds the two frequency points 
(f
1
 and f
2
) at each side of f
0
 at which the signal amplitude is X dB below the 
maximum amplitude. The emission bandwidth is f
2
-f
1
. 
Measurement Parameters: average number, average mode, max hold, span 
and X dB.  
1.  Avg Num 
Specify the number of averages used when calculating the measurement 
result and the default is “Off”. You can use the numeric keys, knob or direction 
keys to modify this parameter. 