Information
RIGOL
DSA800 Options and Accessories
4
VSWR Measurement Kit
DSA800-VSWR option (used together with the VB1020 and DSA800-TG options)
provides measurement functions of S11-related specifications (such as the return
loss, reflection coefficient and VSWR).
Press Meas at the front panel and then press VSWR to enable the VSWR
measurement function. The screen is divided into two windows with the upper
window (the basic measurement window) displaying the sweep trace and the lower
window displaying the measurement wizard and measurement results. Perform two
measurements respectively according to the measurement wizard in the lower
window: measurement with the device under test disconnected (Meas Setup
Cal Open, represented by trace 2) and measurement with the device under test
connected (Meas Setup VSWR, represented by trace 1). The return loss is
determined by the difference (represented by the math trace) of the results of the
two measurements and the reflection coefficient and VSWR are determined by the
return loss.
VB1020
Device under Test