Information
RIGOL 
                DSA800 Options and Accessories 
4 
VSWR Measurement Kit 
DSA800-VSWR  option  (used  together  with  the  VB1020  and  DSA800-TG  options) 
provides  measurement functions of S11-related  specifications (such as  the return 
loss, reflection coefficient and VSWR).   
Press  Meas  at  the front  panel  and  then press  VSWR to enable the VSWR 
measurement  function.  The  screen  is  divided  into  two  windows  with  the  upper 
window (the basic measurement window) displaying the sweep trace and the lower 
window displaying the measurement wizard and measurement results. Perform two 
measurements  respectively  according  to  the  measurement  wizard  in  the  lower 
window: measurement with the device under test disconnected (Meas  Setup   
Cal  Open, represented by  trace 2) and measurement with the device under test 
connected (Meas  Setup  VSWR,  represented by trace 1).  The return loss is 
determined by the difference (represented by the math trace) of the results of the 
two measurements and the reflection coefficient and VSWR are determined by the 
return loss.   
VB1020 
Device under Test       