Data Sheet

Level Display
Logarithmic level axis 1 dB to 200 dB
Linear level axis 0 to reference level
Number of display points 601
Number of traces 3 + math trace
Trace detectors
normal, positive-peak, negative-peak, sample, RMS, voltage average
quasi-peak (with EMI-DSA800 option)
Trace functions clear write, max hold, min hold, average, view, blank
Units of level axis dBm,dBmV, dBμV, nV, μV, mV, V, nW, μW, mW, W
Frequency Response
DSA815 DSA832 DSA875
Frequency response f
c
≥100 kHz, attenuation = 10 dB, relative to 50 MHz, 20ºC to 30 ºC
PA off
100 kHz to 1.5 GHz <0.7 dB
<0.5 dB, <0.3 dB (typ.)
1.5 GHz to 3.2 GHz
3.2 GHz to 7.5 GHz <0.7 dB, <0.3 dB (typ.)
f
c
≥ 1 MHz, attenuation = 10 dB, relative to 50 MHz, 20ºC to 30 ºC
PA on
100 kHz to 1.5 GHz <1.0 dB
<0.7 dB, <0.3 dB (typ.)
1.5 GHz to 3.2 GHz
3.2 GHz to 7.5 GHz <0.9 dB, <0.3 dB (typ.)
Input Attenuation Switching Uncertainty
DSA815 DSA832 DSA875
Setting range 0 to 30 dB, in 1 dB step
Switching uncertainty
f
c
=50 MHz, relative to 10 dB, 20 ºC to 30 ºC
<0.5 dB <0.3 dB
7