- DSA800 Series Spectrum Analyzer ● All-Digital IF Technology ● Frequency Range from 9 kHz up to 7.5 GHz ● Min. -161 dBm Displayed Average Noise Level (Typ.) ● Min. < -98 dBc/Hz @ 10 kHz Offset Phase Noise ● Level Measurement Uncertainty < 0.8 dB ● 10 Hz Minimum Resolution Bandwidth (DSA832/875) ● Up to 7.5 GHz Tracking Generator (DSA8XX-TG) ● Optional Preamplifier (DSA832/875) ● Advanced Measurement Functions (Opt.) ● EMI Filter & Quasi-Peak Detector Kit (Opt. 
- DSA800 Series Spectrum Analyzer Advanced Measurement Keys Basic Keys USB Port TG Output Marker Keys RF Input Product Dimensions: Width X Height X Depth = 361.6 mm x 178.8 mm x 128 mm Benefits of Rigol's all digital IF design The ability to measure smaller signals: on the basis of this technology, the IF filter enables smaller bandwidth settings, which greatly reduce the displayed average noise level. 
- Features and Benefits Distinguish the two nearby signals clearly with the 10 Hz RBW (DSA832/875) Compare the spectrums with different color trace Readout the spectrum peak values with the peak table function Measure lower level signal with the preamplifier turn on Phase noise < -98 dBc/Hz @10 kHz offset (DSA832/875) EMI kit (EMI filter & Quasi-peak & Pass/Fail) The GUI to control the RF demo kit (Transmitter) directly VSWR measurement RIGOL 2 
- RIGOL Spectrum Analyzer Option and Accessory Harmonic Distoration TOI Occupied Bandwidth Channel Power Time Domain Power Emission Bandwidth Carrier to Noise Ratio Adjacent Channel Power Pass/Fail Advanced Measurement Kit (AMK-DSA800) Rack Mount Kit ( RM-DSA800 ) VSWR Bridge ( VB1020/VB1040/VB1080 ) RF Demo Kit ( TX1000 ) RF Demo Kit ( RX1000 ) RF CATV Kit DSA Utility Kit RF Adaptor Kit RF Attenuator Kit RF Cable Kit High Power Attenuator ( ATT03301H ) DSA PC Software ( Ultra Spectrum ) ( 
- Specifications Specifications are valid under the following conditions: the instrument is within the calibration period, is stored for at least two hours at 0 ℃ to 50 ℃ temperature, and is warmed up for 40 minutes. Unless otherwise noted, the specifications in this manual include the measurement uncertainty. Typical (typ.): characteristic performance, which 80 percent of the measurement results will meet at room temperature (approximately 25℃ ). 
- Residual FM 20℃ to 30℃ , RBW = VBW = 1 kHz DSA815 <50 Hz (nom.) Residual FM DSA832 <20 Hz (nom.) DSA875 DSA832 DSA875 Bandwidths Resolution bandwidth (-3 dB) RBW uncertainty Resolution filter shape factor (60 dB: 3 dB) Video bandwidth (-3 dB) Resolution bandwidth (-6 dB) (EMI-DSA800 option) DSA815 100 Hz to 1 MHz, in 1-3-10 sequence <5% (nom.) 10 Hz to 1 MHz, in 1-3-10 sequence <5 (nom. 
- Displayed Average Noise Level (DANL) Frequency PA off PA on 9 kHz to 100 kHz 100 kHz to 5 MHz 5 MHz to 3.2 GHz 3.2 GHz to 6 GHz 6 GHz to 7.5 GHz 100 kHz to 1 MHz 1 MHz to 5 MHz 5 MHz to 3.2 GHz 3.2 GHz to 6 GHz 6 GHz to 7.5 GHz DSA832 DSA875 attenuation = 0 dB, RBW = VBW = 10 Hz, sample detector, trace average ≥ 50, tracking generator off, 20℃ to 30℃ , input impendence = 50 Ω <-110 dBm (typ.) <-110 dBm (typ.) <-125 dBm, <-128 dBm (typ.) <-125 dBm, <-128 dBm (typ.) <-130 dBm, <-134 dBm (typ. 
- Level Display Logarithmic level axis Linear level axis Number of display points Number of traces Trace detectors Trace functions Units of level axis 1 dB to 200 dB 0 to reference level 601 3 + math trace normal, positive-peak, negative-peak, sample, RMS, voltage average quasi-peak (with EMI-DSA800 option) clear write, max hold, min hold, average, view, blank dBm,dBmV, dBμV, nV, μV, mV, V, nW, μW, mW, W Frequency Response Frequency response 100 kHz to 1.5 GHz PA off 1.5 GHz to 3.2 GHz 3.2 GHz to 7. 
- Absolute Amplitude Uncertainty DSA815 DSA875 DSA832 fc = 50 MHz, peak detector, preamplifier off, attenuation = 10 dB, input signal level = -10 dBm, 20 ºC to 30 ºC <0.4 dB <0.3 dB Uncertainty RBW Switching Uncertainty relative to 1 kHz RBW <0.1 dB Uncertainty Reference Level Range Resolution log scale linear scale -100 dBm to +20 dBm, in 1 dB step 0.01 dB 4 digits Preamplifier Gain 100 kHz to 1.5 GHz 1.5 GHz to 3.2 GHz 3.2 GHz to 7.5 GHz DSA815 (standard) 20 dB (nom. 
- RF Input VSWR VSWR 300 kHz to 1.5 GHz 1.5 GHz to 3.2 GHz 3.2 GHz to 7.5 GHz DSA815 attenuation ≥ 10 dB <1.5 (nom.) DSA832 DSA875 <1.5 (nom.) <1.5 (nom.) <1.8 (nom. 
- Spurious Responses Spurious response Intermediate frequency System related sidebands Input related spurious DSA815 DSA832 DSA875 input terminated 50 Ω, attenuation = 0 dB, 20ºC to 30ºC <-90 dBm[1], <-88 dBm (typ.) <-100 dBm (typ.) <-60 dBc referenced to local oscillators, referenced to A/D conversion, referenced to subharmonic of first LO, referenced to harmonic of first LO <-60 dBc mixer level = -30 dBm <-60 dBc NOTE: [1] Except the internal local oscillator (1820 MHz) and its harmonics. 
- Trigger Functions Trigger Trigger source External trigger level free run, video, external 5 V TTL level Input /Output Front Panel Connectors impedance RF input connector impedance Tracking generator output connector 50 Ω (nom.) N female 50 Ω (nom.) N female Internal/ External Reference frequency output level Internal reference impedance connector frequency input level External reference impedance connector 10 MHz +3 dBm to +10 dBm, +8 dBm (typ.) 50 Ω (nom. 
- General Specifications Display Type Resolution Size Colors TFT LCD 800 x 480 pixels 8 inch 64 k Printer Supported Protocol PictBridge Mass Memory Mass memory flash disk (internal), USB storage device (not supplied) Power Supply Input voltage range, AC AC supply frequency Power consumption 100 V to 240 V (nom.) 45 Hz to 440 Hz 35 W (typ.), max. 
- Ordering Information Model Standard accessories Options Optional accessories Description spectrum analyzer, 9 kHz to 1.5 GHz (with preamplifier) spectrum analyzer, 9 kHz to 3.2 GHz spectrum analyzer, 9 kHz to 7.5 GHz spectrum analyzer, 9 kHz to 1.5 GHz (with preamplifier, with tracking generator, factory installed) spectrum analyzer, 9 kHz to 3.2 GHz (with tracking generator, factory installed) spectrum analyzer, 9 kHz to 7.