Datasheet

Frequency
to 50
uncertainty.
Typical (typ.): characteristic performance, which 80 percent of the measurement results will meet at room temperature (approximately 25
).
This data is not warranted and does not include the measurement uncertainty.
Nominal (nom.): the expected mean or average performance or a designed attribute (such as the 50 Ω connector). This data is not warranted
and is measured at room temperature (approximately 25
).
Measured (meas.): an attribute measured during the design phase which can be compared to the expected performance, such as the
amplitude drift variation with time. This data is not warranted and is measured at room temperature (approximately 25
).
NOTE: All charts in this manual are the measurement results of multiple instruments at room temperature unless otherwise noted.
RIGOL
4
Frequency
DSA705 DSA710
Frequency range 100 kHz to 500 MHz 100 kHz to 1 GHz
Frequency resolution 1 Hz
Internal Reference Frequency
DSA705 DSA710
Reference frequency 10 MHz
Accuracy ±[ (time since last calibration × aging rate) + temperature stability + calibration accuracy]
Initial calibration accuracy <1 ppm
Temperature stability
0
to 50
, reference to 25
<2 ppm
Aging rate <2 ppm/year
Frequency Readout Accuracy
Marker resolution span/ (number of sweep points - 1)
Marker uncertainty
±(frequency indication × reference frequency accuracy + 1% × span + 10% × resolution bandwidth +
marker resolution)
Frequency Counter
Resolution 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz
Uncertainty ±(frequency indication × reference frequency accuracy + counter resolution)
Frequency Span
Range 0 Hz, 100 Hz to maximum frequency of instrument
Uncertainty ±span/ (number of sweep points - 1)
SSB Phase Noise
DSA705 DSA710
20
to 30
, f
c
= 500 MHz 20
to 30
, f
c
= 1 GHz
Carrier offset
10 kHz <-80 dBc/Hz