Instructions
RIGOL  Chapter 2 Command System 
2-42    MSO7000/DS7000 Programming Guide 
:CHANnel<n>:SCALe 
Syntax 
:CHANnel<n>:SCALe <scale> 
:CHANnel<n>:SCALe? 
Description 
Sets or queries the vertical scale of the specified channel. The default unit is V. 
Parameter 
Name 
Type 
Range 
Default 
<n> 
Discrete 
{1|2|3|4} 
—— 
<scale> 
Real 
Related to input impedance and probe 
ratio: 
when the input impedance is 1 MΩ, the 
probe ratio is 1X: 1 mV to 10 V 
when the input impedance is 50 Ω, the 
probe ratio is 1X: 1 mV to 1 V 
100 mV 
Remarks 
You can use the :CHANnel<n>:VERNier command to enable or disable the fine 
adjustment setting for the vertical scale of the specified channel. By default, the fine 
adjustment is Off. At this time, you can set the vertical scale at 1-2-5 step, i.g. 10 mV, 20 
mV, 50 mV, 100 mV……100 V (probe ratio is 10X). When the fine adjustment is enabled, 
you can further adjust the vertical scale within a relatively smaller range to improve 
vertical resolution. If the amplitude of the input waveform is a little bit greater than the 
full scale under the current scale and the amplitude would be a little bit lower if the next 
scale is used, fine adjustment can be used to improve the amplitude of waveform display 
to view signal details. 
Return 
Format 
The query returns the vertical scale in scientific notation. 
Example 
:CHANnel1:SCALe 1      /*Sets the vertical scale of CH1 to 1 V.*/ 
:CHANnel1:SCALe?      /*The query returns 1.*/ 
:CHANnel<n>:IMPedance 
Syntax 
:CHANnel<n>:IMPedance <impedance> 
:CHANnel<n>:IMPedance? 
Description 
Sets or queries the input impedance of the specified analog channel. The default unit is 
Ω. 
Parameter 
Name 
Type 
Range 
Default 
<n> 
Discrete 
{1|2|3|4} 
—— 
<impedance> 
Discrete 
{OMEG|FIFTy} 
OMEG 
Remarks 
  OMEG: 1 MΩ. This indicates that the input impedance of the oscilloscope is rather 
high, and the current flowing from the circuit under test to the oscilloscope can be 
ignored. 
  FIFTy: 50 Ω. This indicates that the oscilloscope shall be matched with a device 
whose output impedance is 50 Ω. 
Return 
Format 
The query returns OMEG or FIFT. 
Example 
:CHANnel1:IMPedance FIFTy      /*Sets the input impedance of CH1 to FIFTy.*/ 
:CHANnel1:IMPedance?              /*The query returns FIFT.*/ 