Instructions
RIGOL  Chapter 2 Command System 
2-40    MSO7000/DS7000 Programming Guide 
:CHANnel<n>:COUPling 
Syntax 
:CHANnel<n>:COUPling <coupling> 
:CHANnel<n>:COUPling? 
Description 
Sets or queries the coupling mode of the specified channel. 
Parameter 
Name 
Type 
Range 
Default 
<n> 
Discrete 
{1|2|3|4} 
—— 
<coupling> 
Discrete 
{AC|DC} 
DC 
Remarks 
  AC: the DC components of the signal under test are blocked. 
  DC: both DC and AC components of the signal under test can pass through the 
channel. 
Return 
Format 
The query returns AC or DC. 
Example 
:CHANnel1:COUPling AC        /*Selects the AC coupling mode.*/ 
:CHANnel1:COUPling?            /*The query returns AC.*/ 
:CHANnel<n>:DISPlay 
Syntax 
:CHANnel<n>:DISPlay <bool> 
:CHANnel<n>:DISPlay? 
Description 
Turns on or off the specified channel; or queries the on/off status of the specified 
channel. 
Parameter 
Name 
Type 
Range 
Default 
<n> 
Discrete 
{1|2|3|4} 
—— 
<bool> 
Bool 
{{1|ON}|{0|OFF}} 
0|OFF 
Return 
Format 
The query returns 1 or 0. 
Example 
:CHANnel1:DISPlay ON        /*Enables CH1.*/ 
:CHANnel1:DISPlay?            /*The query returns 1.*/ 
:CHANnel<n>:INVert 
Syntax 
:CHANnel<n>:INVert <bool> 
:CHANnel<n>:INVert? 
Description 
Turns on or off the waveform invert for the specified channel; or queries the on/off 
status of the waveform invert for the specified channel. 
Parameter 
Name 
Type 
Range 
Default 
<n> 
Discrete 
{1|2|3|4} 
—— 
<bool> 
Bool 
{{1|ON}|{0|OFF}} 
0|OFF 
Remarks 
When the waveform invert is turned off, the waveform is displayed normally; when 
the waveform invert is turned on, the voltage values of the displayed waveform are 
inverted. 
Return Format 
The query returns 1 or 0. 