Instructions
Chapter 2 Command System  RIGOL 
MSO7000/DS7000 Programming Guide  2-109   
:MEASure:CLEar 
Syntax 
:MEASure:CLEar <item> 
Description 
Clears any one or all of the 10 measurement items that have been turned on. 
Parameter 
Name 
Type 
Range 
Default 
<item> 
Discrete 
{ITEM1|ITEM2|ITEM3|ITEM4|ITEM5|ITEM6|I
TEM7|ITEM8|ITEM9|ITEM10|ALL} 
—— 
Remarks 
Run the :MEASure:ITEM command to open one of the 41 parameters that is required to 
be measured. The 10 measurement items turned on last time are determined by the 
order in which you turned them on, and they will not be affected if you delete one or 
multiple measurement items. 
Example 
:MEASure:CLEar ITEM1        /*Clears a measurement item ITEM1.*/ 
:MEASure:THReshold:SOURce 
Syntax 
:MEASure:THReshold:SOURce <source> 
Description 
Sets the threshold source. 
Parameter 
Name 
Type 
Range 
Default 
<source> 
Discrete 
{CHANnel1|CHANnel2|CHANnel3|CHANnel4|
MATH1|MATH2|MATH3|MATH4} 
CHANnel1 
Remarks 
Modifying the threshold will affect the measurement results of time, delay, and phase 
parameters. 
Example 
:MEASure:THReshold:SOURce CHANnel2 /*Sets the threshold source to CH2.*/ 
:MEASure:THReshold:DEFault 
Syntax 
:MEASure:THReshold:DEFault 
Description 
Sets the threshold level of the analog channel in auto measurement to a default value. 
:MEASure:MODE 
Syntax 
:MEASure:MODE <mode> 
:MEASure:MODE? 
Description 
Sets or queries the measurement mode. 
Parameter 
Name 
Type 
Range 
Default 
<mode> 
Discrete 
{NORMal|PRECision} 
NORMal 
Remarks 
  NORMal: executes measurement of up to 1 Mpts. 
  PRECision: executes measurement of up to 500 Mpts, improving the resolution of 
measurement results. Note, in this mode, the refresh rate of the waveforms may be 
reduced. 
Example 
The query returns NORM or PREC. 
:MEASure:MODE PRECision            /*Sets the measurement mode to PRECision.*/ 
:MEASure:MODE?                            /*The query returns PREC.*/ 