Instructions
RIGOL  Chapter 2 To Set the Vertical System 
2-10  MSO7000/DS7000 User's Guide 
Amplitude Unit 
Select the amplitude display unit for the current channel. The available units are W, A, 
V, and U. When the unit is changed, the unit related to the relevant functions of the 
channel will change accordingly. 
Press 1  More  Unit, press Unit continuously or rotate the multifunction knob 
  to select a unit. You can also enable the touch screen to select the desired unit 
with touch gestures. The default unit for the value is V. 
Channel Delay 
When using an oscilloscope for actual measurement, the transmission delay of the 
probe cable may bring relatively greater errors (zero offset). MSO7000/DS7000 
allows you to set a delay time for calibrating the zero offset of the corresponding 
channel. Zero offset is defined as the offset of the crossing point of the waveform 
and trigger level line relative to the trigger position, as shown in Figure 2-2. 
Figure 2-2 Zero Offset 
Press 1  More  Ch-Ch Skew, rotate the multifunction knob    or use the 
numeric keypad to set the desired delay calibration time. The available range of the 
delay calibration time is from -100 ns to 100 ns.   
Note: This parameter is related to the instrument model and the current horizontal 
time base setting. The larger the horizontal time base is, the larger the setting 
step will be.   
Actual Trigger Position 
Trigger 
Level 