Instructions
Chapter 2 To Set the Vertical System    RIGOL 
MSO7000/DS7000 User's Guide  2-9 
Passive Probe 
If a passive probe is connected, for example, when RIGOL's RP3500A probe model 
is connected to the oscilloscope, press Probe to open the probe menu. Its 
sub-menus are grayed out and disabled. 
Active Probe 
If an active probe is connected, for example, when RIGOL's RP7150 probe model is 
connected to the oscilloscope, press Probe to open the probe menu, and you can 
operate on the menu item. 
  Probe Delay 
To avoid measurement result errors arising from the transmission delay of the 
probe cable, the oscilloscope provides the probe delay adjustment function for 
the active probe. Press Skew,  and then rotate the multifunction knob    or 
use the numeric keypad to set the probe delay time. 
  Bias Voltage 
The oscilloscope provides the bias voltage adjustment function for active probes. 
This function is used to adjust the signal under test that exceeds the input 
dynamic range of the probe amplifier to an appropriate range to ensure the 
integrity of the signal under test. Press Bias, rotate the multifunction knob   
or use the numeric keypad to set the DC bias voltage.   
  Probe Information 
Press About to view information about the probe currently connected, such as 
the manufacturer, model, serial number, and the last calibration date. 
  Probe Calibration 
Press Calibration to start zero calibration automatically for the probe.   
Tip   
For some of the ordinary passive probes, the oscilloscope can recognize their 
probe ratio automatically. If not recognized, please refer to "Probe Ratio" to set 
the probe ratio.   
Tip   
  When a 50 Ω active probe is connected, the input impedance (refer to "Input 
Impedance") of the channel will be automatically set to "50 Ω". 
  For the active probe, the oscilloscope can recognize the probe ratio 
automatically. If the currently connected probe supports a variety of probe 
ratios, please refer to "Probe Ratio" to set the probe ratio. 










