Instructions
RIGOL  Chapter 2 To Set the Vertical System 
2-8  MSO7000/DS7000 User's Guide 
Waveform Invert 
Press 1 to open the setting menu of CH1. Then, press Invert continuously to enable 
or disable the waveform invert. When enabled, the channel label is displayed, as 
shown in the following figure. 
When disabled, the waveform is displayed normally; when enabled, the voltage 
values of the displayed waveform are inverted (as shown in Figure 2-1). Enabling the 
waveform invert will also change the result of math function and waveform 
measurement. 
(a) "Invert" Off 
(b) "Invert" On 
Figure 2-1 Waveform Invert 
Note: When the waveform invert is enabled, the trigger edge or the trigger polarity 
will change (e.g. Edge trigger, Pulse trigger, or Slope trigger). 
To Set the Probe 
The analog channel of this oscilloscope not only supports the common passive probe, 
but also the active probe. It can automatically recognize the currently connected 
probe type and its probe ratio. For detailed technical information of the probes, 
please refer to the corresponding Probe User’s Guide. 
Press 1 to open the setting menu of CH1. Then, press Probe to open the probe 
menu. If different probes are connected to the oscilloscope, different information 
about the probe will be displayed, and you can perform the different operations on 
the menu. 