Instructions
Chapter 11 Protocol Decoding    RIGOL 
MSO7000/DS7000 User's Guide  11-41 
time, you can view the details in the "Packets" view.   
  Export: If you select "Packets", you can export time and the 
corresponding decoded data. Press Export, the save setting menu is 
displayed. You can export the data table of the packets to the internal 
or external USB storage device (when detected) in CSV format. For 
detailed saving operation, refer to descriptions in "Store and Load". 
  Jump to: Set the operating status of the oscilloscope to "STOP". Then, 
select the data of the specified time from the event table. Press Jump 
to and the specified data waveforms are displayed in the center of the 
screen. 
  Decode: The setting of this menu enables you to quickly view the 
corresponding event tables of four decode modules (Decode1, 
Decode2, Decode3, and Decode4). Press Decode and rotate the 
multifunction knob    to select a decode module, then press down 
the knob to select it. You can also press Decode continuously or 
enable the touch screen to select it. 
1553B Decoding (Option) 
The oscilloscope samples the 1553B signal, and judges each data point to be logic 
"1" or logic "0" according to the preset threshold level. 1553B decoding is required to 
specify the data channel source and the threshold. 
In the decode setting menu, press Decode1  Bus Type, then rotate the 
multifunction knob    to select "1553B". Press down the knob to select it. You can 
also press Bus Type continuously or enable the touch screen to select it. 
1.  Enable or disable the bus 
Press Bus Status continuously to enable or disable the decoding function. 
2.  Quickly apply 1553B trigger settings to 1553B decoding 
Press Copy Trig to copy the 1553B trigger settings and apply them to 1553B 
decoding function (set the corresponding I2S decoding parameters 
automatically).   
3.  Set the data channel source and the threshold 
  Press Data and rotate the multifunction knob    to select the desired data 
channel source, and then press down the knob to select it. You can also 
press Data continuously or enable the touch screen to select it. The 
available channels include CH1-CH4. 
  Press Threshold, then rotate the multifunction knob    or use the pop-up 
numeric keypad to set the threshold of the source channel. When you 
modify the threshold of the channel, a dotted line displaying the current 
threshold level is displayed on the screen. The dotted line disappears in 