Instructions
Chapter 11 Protocol Decoding    RIGOL 
MSO7000/DS7000 User's Guide  11-37 
I2S Decoding (Option) 
The oscilloscope samples the I2S signal and judges each data point as logic "1" or 
logic "0" according to the preset threshold level. I2S decoding is required to specify 
the serial clock, channel signal, and the data's source channel. You need to set 
Alignment, WS Low, and other parameters. 
In the decode setting menu, press Decode1  Bus Type, then rotate the 
multifunction knob    to select "I2S". Press down the knob to select it. You can also 
press Bus Type continuously or enable the touch screen to select it. 
1.  Enable or disable the bus 
Press Bus Status continuously to enable or disable the decoding function. 
2.  Quickly apply I2S trigger settings to I2S decoding 
Press Copy Trig to copy the I2S trigger settings and apply them to I2S 
decoding function (set the corresponding I2S decoding parameters 
automatically). For analog channels, the threshold level settings will also be 
copied. 
3.  Source Setting 
Press Sources to enter the source setting menu. 
  Set the serial clock channel source, threshold, and clock edge 
  Press SCLK and rotate the multifunction knob    to select the desired 
channel, and then press down the knob to select it. You can also press 
SCLK continuously or enable the touch screen to select it. The 
available channels include CH1-CH4 and D0-D15. 
  When you select the analog channel (CH1-CH4), press SCLK Thre 
below SCLK, then rotate the multifunction knob    or use the pop-up 
numeric keypad to set the threshold of the SCLK. When you modify the 
threshold of the clock channel, a dotted line displaying the current 
threshold level is displayed on the screen. The dotted line disappears in 
about 2 s after you stop modifying the threshold. 
  Press SCLK Edge continuously to select "Rising( )" or 
"Falling( )" as the desired clock edge. 
     TSS    Sync Frame Frame    PL    HCRC      CYC Data                       TCRC   