Instructions
Chapter 6 Operations and Measurements    RIGOL 
MSO7000/DS7000 User's Guide  6-37 
  X Cursor 
X  cursor  is  a  vertical  solid/dotted  line  that  is  used  to  make  horizontal 
adjustments. It can be used to measure time (s), frequency (Hz), phase (°), and 
ratio (%). 
  Cursor A is a vertical solid line, and Cursor B is a vertical dotted line. 
  In the  X-Y cursor  mode, X cursor is used  to  measure the waveform 
amplitude of CH1. 
  Y Cursor 
Y cursor is a horizontal solid/dotted line that is used to make vertical adjustment. 
It can be used to measure amplitude (the unit is the same as that of the source 
channel amplitude) and ratio (%). 
  Cursor A is a horizontal solid line, and Cursor B is a horizontal dotted line. 
  In the  X-Y  cursor  mode, Y cursor  is used  to  measure  the  waveform 
amplitude of CH2. 
Press Cursor on the front panel to open the cursor measurement menu, or enable 
the  touch  screen  and  then  tap  the  function  navigation  icon    at  the  lower-left 
corner of the screen to open the function navigation. Then, tap the "Cursor" icon to 
open the cursor measurement menu. 
In the cursor measurement menu, press Mode and then rotate the multifunction 
knob    to select the desired cursor mode (the default is "OFF"). Press down the 
knob to select it. You can also press the Mode key continuously or enable the touch 
screen to select the desired mode. The available modes are "Manual", "Track", and 
"XY". When you select "OFF", the cursor measurement function is disabled.   
Note: When the Time Base Mode is set to "XY", the XY cursor measurement mode is 
valid. 
Y Cursor (Cursor B) 
Y Cursor (Cursor A) 
X Cursor (Cursor A) 
X Cursor (Cursor B) 